Fluidigm Polaris System Teardown: Illuminator, Camera, and Optics
0:00 Optical System Overview: Disassembly reveals heavy lenses (~4kg each), focusing on alignment and positioning rather than magnification.
0:49 Filter Wheel Assembly: Features five narrow band filters (488nm, 525nm, 570nm, 630nm, 700nm) for fluorescence detection and excitation light rejection in biochemical processes.
2:01 Lumen Dynamics Light Source: A self-contained unit with USB and TTL interfaces, powering individual LED drivers for each wavelength (430nm, 475nm, 520nm, 575nm, 630nm). A 385nm LED position is unpopulated.
2:50 Custom Design: The light source PCB is application-specific, with wavelengths labeled directly on the board.
3:57 Simple TTL Control: LEDs are controlled via a straightforward 5V TTL signal, with potential for additional unused control lines.
4:48 Dichroic Mirrors & Light Path: White light test reveals the path of each wavelength through the system, highlighting the use of dichroic mirrors for color separation and combination.
6:36 Unique 575nm Yellow LED: Achieved by filtering a white LED, unlike other wavelengths using direct color LEDs. This approach is likely due to the difficulty in obtaining a pure yellow LED at high intensities.
7:51 Light Guide for Uniform Illumination: A square, likely fiber optic rod ensures even light distribution across the sample. A feedback loop involving a looped-back fiber and sensor compensates for light loss.
8:49 Trapezoidal Light Guides: Shape of the light guides provides keystone correction for rectangular illumination of the sample.
9:56 LED Package Variations: The 520nm green LED uses a phosphor, likely for precise wavelength tuning. Other LEDs appear to be direct-color emitters. The white LED appears to be a standard Phloxlight module.
11:48 Runtime Tracking: EEPROM data reveals only 20 minutes of runtime for the green LED, indicating minimal overall system use.
12:55 Peltier-Cooled Camera: The camera system includes a 50-megapixel Kodak KAI-50100 CCD with a Peltier cooler for noise reduction.
13:51 Evidence of Troubleshooting/Modifications ("Bodges"): Extra decoupling capacitors and other modifications suggest attempts to address noise or other issues.
14:54 Dual-Output CCD Architecture: The CCD outputs two parallel data streams to manage high bandwidth requirements.
17:01 Camera Testing & Software: Basic functionality tested using pinhole lens and Windows 7 drivers. Smearing and potential dynamic range limitations observed.
17:17 Software Features: Software includes options for clock frequency, gain, exposure control, and triggering, but PFG function remains unknown.
18:40 Availability on eBay: Both the illuminator and the camera will be listed on eBay.
19:59 Temperature Control Issue: While the Peltier cooler's target temperature can be set, there's no readout for the actual achieved temperature.